當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 158Sα-蒎烯C10H16氣體檢測管158S α-PINENE 20-300ppm
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
α-蒎烯C10H16氣體檢測管158S α-PINENE 20-300ppm
Tube 58S
1. PERFORMANCE
1) Measuring range :5-300 ppm 2.5-150 ppm
Number of pump strokes 1(100ml) 2(200ml)
2) Sampling time :1 minute/1 pump stroke
3) Detectable limit :0.5 ppm(200ml)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :White→Yellow
2. RELATIVE STANDARD DEVIATION
RSD-low:15% RSD-mid.:15% RSD-high : 10%
3. CHEMICAL REACTION
A polymer of Styrene is produced by sulphuric acid.
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
300
250
200
150
100
50
20
5
ppm
GAS To pump
158S STYRENE
Substance ppm Interference ppm Coexistence
Acrylonitrile FIG.1 Tafhfeec atecdc.uracy of readings is not 400 Lower readings are given.
Butadiene 3 Similar stain is produced. 5 Uneven discolouration is produced
FIG.2 and higher readings are given.
Formaldehyde 〃15 Yellowish orange stain is produced
FIG.3 and higher readings are given.
Acetaldehyde 〃350 Similar stain is produced and
higher readings are given.
Methyl alcohol The accuracy of readings is not 0.35% Pale discolouration is produced
affected. and higher readings are given.
Ethyl alcohol 〃0.18% 〃
Ethyl acetate 〃700 〃
Butyl acetate 〃700 〃
(NOTE)
In case of 2 pump strokes, following formula is available for actual concentration.
Actual concentration=1/2×Reading value.
0
200
100
Acrylonitrile Butadiene
FIG.1 Influence of Acrylonitrile Styrene indicating value 〔
ppm〕
500 1000 0
200
100
FIG.2 Influence of Butadiene
Styrene indicating value 〔ppm〕
〔ppm〕? 50 100 0
200
100
Formaldehyde
FIG.3 Influence of Formaldehyde
Styrene indicating value 〔ppm〕
〔ppm〕? 100 200〔ppm〕?
Coexistence
Interference
CH:?CH2 CH–CH2
H2SO4+?
請輸入賬號
請輸入密碼
請輸驗證碼
以上信息由企業自行提供,信息內容的真實性、準確性和合法性由相關企業負責,化工儀器網對此不承擔任何保證責任。
溫馨提示:為規避購買風險,建議您在購買產品前務必確認供應商資質及產品質量。