當前位置:上海境宣實業有限公司>>日本光明北川式氣體檢測管>>氣體快速檢測管>> 120SB硫化氫H2S氣體檢測管120SB HYDROGEN SULPHIDE 0.75-300ppm
室內空氣甲醛檢測管No.713 FORMALDEHYDE 0.01-0
氯乙烯CH2=CHCl氣體測毒管132SC VINYL CHLORID
氯乙烯CH2=CHCl氣體檢知管132SB VINYL CHLORID
硫化氫H2S氣體檢測管120SB HYDROGEN SULPHIDE 0.75-300ppm
Tube 20SB
1. PERFORMANCE
1) Measuring range :3-150 ppm 1-50 ppm 0.75-37.5 ppm 6-300 ppm
Number of pump strokes 1(100ml) 3(300ml) 4(400ml) 1/2(50ml)
2) Sampling time :1 minutes/1 pump stroke
3) Detectable limit :0.3 ppm(300ml)
4) Shelf life :3 years
5) Operating temperature :0~40℃
6) Reading :Direct reading from the scale calibrated by 1 pump stroke
7) Colour change :White→Dark brown
2. RELATIVE STANDARD DEVIATION
RSD-low:10% RSD-mid.:5% RSD-high:5%
3. CHEMICAL REACTION
By reacting with Lead acetate(II), Lead sulphide is produced.
H2S+Pb(CH3CO2)2→PbS+2CH3CO2H
4. CALIBRATION OF THE TUBE
PERMEATION TUBE METHOD
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
150
140
130
120
110
100
90
80
70
60
50
40
30
20
10
6 3
100ml
ppm
??
????
120SB HYDROGEN SULPHIDE
Substance Interference ppm Coexistence
Sulphur dioxide FIG.1 Tish neo at cacffuercatceyd .of readings 12 Higher readings are given.
Mercaptans FIG.2 〃550 〃
Nitrogen dioxide FIG.3 〃2 Lower readings are given.
(NOTE)
In case of 1 / 2, 2, 3 or 4 pump strokes, following formula is available for the actual concentration.
Actual concentration=Reading value ×
0
60
80
20
Sulphur dioxide〔ppm〕?
FIG.1 Influence of Sulphur dioxide
Hydrogen sulphide indicating value 〔ppm〕
200 400
300ml
0
60
80
20
Ethyl mercaptan〔ppm〕?
FIG.2 Influence of Ethyl mercaptan
Hydrogen sulphide indicating value 〔ppm〕
200 400 600
300ml
0
100
150
50
Nitrogen dioxide〔ppm〕?
FIG.3 Influence of Nitrogen dioxide