Tube 11U
1. PERFORMANCE
1) Measuring range :10-1,000 ppm
Number of pump strokes 1(100mL)
2) Sampling time :1.5 minutes / 1 pump stroke
3) Detectable limit : 5 ppm
4) Shelf life :2 years
5) Operating temperature :10~40℃
6) Temperature compensation :Necessary (See "TEMPERATURE CORRECTION TABLE" )
7) Reading :Direct reading from the scale calibrated by 1 pump stroke
8) Colour change :Yellow→Brown
2. RELATIVE STANDARD DEVIATION
RSD-low :10 % RSD-mid.:10 % RSD-high :5%
3. CHEMICAL REACTION
Chromium oxide is reduced.
CH 3CO 2C 2H 5+Cr 6++H 2SO 4→Cr 3+
4. CALIBRATION OF THE TUBE
GAS CHROMATOGRAPHY
5. INTERFERENCE AND CROSS SENSITIVITY
GAS To pump
111U ETHYL ACETATE
TEMPERATURE CORRECTION COEFFICIENT TABLE
Scale Correction Coefficient (at 20℃)
Readings 10 ℃ 15 ℃ 20 ℃ 25 ℃ 30 ℃ 35 ℃ 40℃
(ppm) (50 °F) (59°F) (68°F) (77 °F) (86°F) (95°F) (104°F )
1,000 1.33 1.17 1.0 0.87 0.74 0.64 0.53
Substance Interference Coexistence
Alcohols Similar stain is produced. Higher reading are given.
Esters 〃〃
Ketones 〃〃
Aromatic hydrocarbons 〃〃
Whole reagent is discoloured
It the top of brown stain is
Paraffin hydrocarbons clear, the accuracy of
Pale brown.
readings is not affected.
Halogenated hydrocarbons
〃〃
FIG.1
0
200
400
600
800
1000
Trichloroethylene 〔ppm〕
FIG.1 Influence of Trichloroethylene
E thyl acetate indicating value 〔ppm 〕
100 200 300
800 1.38 1.19 1.0 0.86 0.73 0.63 0.53
600 1.40 1.20 1.0 0.86 0.72 0.63 0.53
400 1.40 1.20 1.0 0.85 0.70 0.58 0.46
200 1.40 1.20 1.0 0.84 0.68 0.55 0.42
100 1.50 1.25 1.0 0.81 0.62 0.48 0.33
50 1.50 1.25 1.0 0.77 0.54 0.43 0.32
30 1.50 1.25 1.0 0.77 0.53 0.42 0.30
10 1.50 1.25 1.0 0.75 0.50 0.40 0.30