700B Headstage Calibration
The steps below provide a quick check of the calibration of the MultiClamp 700B. It is assumed that appropriate shielding (as described in “Test the Noise”, above) will be used during these tests.
1. Connect an oscilloscope to the front panel PRIMARY or SCOPE Output BNC.
2. Synchronize the oscilloscope by connecting to the rear panel SYNC OUTPUT BNC. Press the “Reset to Program Defaults” button in the MultiClamp 700B Commander to standardize the MultiClamp 700B.
50 G Range
1. Press the “Options” button, choose the Gains tab, and select the 50 G feedback resistor in the Voltage Clamp pane. Return to the main MultiClamp 700B Commander window.
2. Plug the PATCH connector of the PATCH-1U model cell into the CV-7 headstage.
3. Set Seal Test amplitude to 100 mV and frequency to 50 Hz, then check the box to make it active.
4. Press Auto Cp Fast to remove the bulk of electrode capacitance transient.
5. The resulting waveform should be square, except for an initial overshoot (possibly twice the size of the steady-state response) that settles to the baseline in about 1 to 2 ms. The rise time to the peak of the overshoot should be about 50 μs. The steady-state amplitude following the transient should be ~500 mVp-p (±50 mV).
5 G Range
1. Change the feedback resistor to 5 G in the Options / Gains menu.
2. Press Auto Cp Fast.
3. The step response should be ~50 mVp-p (±5 mV).
500 M Range
1. Press the “Reset to Program Defaults” button. (By default, the 500 M
range is selected.)
2. Check Seal Test and set the amplitude to 25 mV.
3. Plug the CELL connector of the PATCH-1U model cell into the CV-7 headstage.
4. Press the Auto Whole Cell button.
5. Press Auto Cp Fast button.
6. The step response should be ~25 mVp-p.
50 M Range
1. Change the feedback resistor to 50 M.
2. Increase Output Gain to 10.
3. Press Auto Whole Cell and Auto Cp Fast.
4. The step response should be ~25 mVp-p.
相關產品
免責聲明
- 凡本網注明“來源:化工儀器網”的所有作品,均為浙江興旺寶明通網絡有限公司-化工儀器網合法擁有版權或有權使用的作品,未經本網授權不得轉載、摘編或利用其它方式使用上述作品。已經本網授權使用作品的,應在授權范圍內使用,并注明“來源:化工儀器網”。違反上述聲明者,本網將追究其相關法律責任。
- 本網轉載并注明自其他來源(非化工儀器網)的作品,目的在于傳遞更多信息,并不代表本網贊同其觀點和對其真實性負責,不承擔此類作品侵權行為的直接責任及連帶責任。其他媒體、網站或個人從本網轉載時,必須保留本網注明的作品第一來源,并自負版權等法律責任。
- 如涉及作品內容、版權等問題,請在作品發表之日起一周內與本網聯系,否則視為放棄相關權利。