SC-6536Z automatic distillation measuring instrument
Instrument Overview
This instrument is the third generation product newly developed by our company. It is designed and manufactured according to the standard test method of the national standard GB/T6536-2010 "Petroleum Products - Determination of Atmospheric Distillation Characteristics", and can measure two samples simultaneously. It can meet the national standard GB/T3146-2010 "Determination of Distillation Range of Industrial Aromatic Hydrocarbons and Related Materials - Part 1: Distillation Method" and international standards such as ASTM D86/ASTM D850/ASTM D1078-IP195/ISO 3405. It integrates mechanical, optical, and electronic technology, using imported sensors, and the level reading of the graduated cylinder adopts an imported CNC optical tracking and detection system. Automatically complete the entire distillation process experiment, widely applicable to the measurement and analysis of distillation characteristics of distillate fuels such as natural gasoline (stable light hydrocarbons), light and intermediate fractions, automotive spark ignition engine fuels, aviation gasoline, jet fuel, diesel and kerosene, as well as naphtha and naphtha solvent oil products under normal pressure.
Technical Parameters
1. Working power supply: AC220V ± 10% 50Hz 2.6KW
2. Operation mode: Windows operating system 10.4 inch color LCD touch screen
3. Test samples: 2 (two identical or different samples can be tested simultaneously)
4. Temperature range: 0~450 ℃, resolution 0.1 ℃, imported from Germany PT100 temperature sensor
5. Refrigeration method: German imported compressor refrigeration
6. Distillation rate: 2-10 mL/min (freely set, automatically adjusted)
7. Volume detection range: 0-100mL, resolution 0.1mL
8. Cold bath temperature range: 0~80 ℃, temperature control accuracy: 0.2 ℃
9. Cold trap temperature range: 0~60 ℃, temperature control accuracy: 0.2 ℃
10. Air pressure measurement range: 300~1100hpa, accuracy ± 3hpa, built-in pressure sensor, automatic value correction
11. Safety protection system: built-in ultraviolet flame sensor automatically monitors, and automatically opens the protective gas valve when a flame occurs
12. Protective gas interface: φ 7.5~8mm; The protective gas is nitrogen or carbon dioxide, and the pressure is not less than 0.6Mpa
13. Overall dimensions of the instrument: 720X500X660cm (length/width/height), net weight: about 90kg
14. Operating ambient temperature: 5~40 ℃ Operating relative humidity: ≤ 80%
Performance characteristics
1. The temperature of the cold bath and the cold trap are controlled by a segmented program. The cold bath part adopts a metal bath technology that integrates condensing tubes and refrigeration evaporation tubes, ensuring direct conduction of cold and heat without liquid heat transfer media, which is both safe and convenient
2. The instrument adopts a lifting and self-locking lifting control technology, which allows the heating furnace to quickly lift and fall back, allowing it to stay at any position, truly realizing the technology of non polarity adjustment
3. The heating furnace adopts a ceramic heating device pulse modulated infrared radiation heating method, with fast heating and cooling speed, ensuring safety and reliability
4. Compressors, heating ceramic components, temperature sensors, and refrigeration diverter valves are all imported components, ensuring the accuracy and reliability of instrument data, while maintaining a long service life and low failure rate
5. The test data can be stored without restriction and viewed at any time. The built-in micro thermal printer outputs the test results, and can also be exported through a USB flash drive to print data reports on the computer;
6. Two samples can be measured at the same time. After the experiment, the furnace frame automatically slides down, cutting off heat and waste heat. The electric furnace cooling fan can automatically start rapid cooling and improve testing efficiency
7. Infrared modulation optical technology realizes liquid level tracking detection and initial distillation detection, without environmental interference, automatically detecting the position of volume zero point and accurately determining the dry point.
8. Coated quartz orifice plate, heat insulation without cracking, unique orifice plate installation structure, replacement of orifice plates with different apertures without adjustment
9. Optional laboratory intelligent control module system can be equipped to achieve remote intelligent information management of the Internet of Things. The cloud platform can monitor the operation status of this instrument in real time, including user information, usage time, test duration, usage frequency, different usage scenarios, and other data.
免責聲明
- 凡本網注明“來源:化工儀器網”的所有作品,均為浙江興旺寶明通網絡有限公司-化工儀器網合法擁有版權或有權使用的作品,未經本網授權不得轉載、摘編或利用其它方式使用上述作品。已經本網授權使用作品的,應在授權范圍內使用,并注明“來源:化工儀器網”。違反上述聲明者,本網將追究其相關法律責任。
- 本網轉載并注明自其他來源(非化工儀器網)的作品,目的在于傳遞更多信息,并不代表本網贊同其觀點和對其真實性負責,不承擔此類作品侵權行為的直接責任及連帶責任。其他媒體、網站或個人從本網轉載時,必須保留本網注明的作品第一來源,并自負版權等法律責任。
- 如涉及作品內容、版權等問題,請在作品發表之日起一周內與本網聯系,否則視為放棄相關權利。