EN149 | EN13274-7 | EN13274-7 | GB 2626 | GB 2626 | 42 CFR 84 | 42 CFR 84 | |
Aerosol | see EN 13274-7 | NaCl | PaO | NaCl | PaO/DOP | NaCl | DOP |
Mean diameter | see EN 13274-7 | 0.06–0.1µm | 0.29–0.45µm | 0.055–0.095 µm | 0.165–0.205 µm | 0.055–0.095µm | 0.165–0.205µm |
Standard deviation | see EN 13274-7 | 2–3 | 1.6–2.2 | <1.86(by additional software module) | <1.6(by additional software module) | <1.86(by additional software module) | <1.6(by additional software module) |
Concentration | see EN 13274-7 | 4–12mg/m³ | 15–25mg/m³ | <200 mg/m³ | (50mg/m³)<200mg/m³ | < 200mg/m³ | <200mg/m³ |
Discharge | – | – | – | required | required | required | required |
Air flow | see EN 13274-7 | 95 l/min | 95 l/min | 85±4 l/min | 85±4 l/min | 85±4 l/min | 85±4.25 l/min |
Temperature | see EN 13274-7 | 22 ± 3 °C | – | 25 ± 5 °C | 25 ± 5 °C | 25 ± 5 °C | 25 ± 5 °C |
Rel. humidity | see EN 13274-7 | < 40 % | – | 20–40 % | – | 20 – 40 % | 20 – 40 % |
(by compressed air) | (by compressed air) | (by compressed air) | |||||
Measurement device | see EN 13274-7 | Sodium flame photometer | Light scattering photometer | particle detector | particle detector | Light scattering photometer | Light scattering photometer |
Measuring time | see EN 13274-7 | 30 s | 30 s | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading |
Pause time | see EN 13274-7 | 180 s | 180 s | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading | lowest eff. during loading |
Exposition | 120 mg | 120 mg | 120 mg | 200 ± 5 mg | 200 ± 5 mg | 200 ± 5 mg | 200 ± 5 mg |
PMFT remarks | O.K. | O.K. | O.K. | O.K. with upgrade KIT | O.K. with upgrade KIT | O.K. with upgrade KIT | O.K. with upgrade KIT |
測量范圍(尺寸) | 0,10 – 40 µm |
體積流量 | 1 – 27 m3/h (Druckbetrieb) |
電源 | 115/230 V, 50/60 Hz |
外型尺寸 | approx. 600 • 1,800 • 900 mm (W • H • D) |
安裝條件 | 19 –23 °C |
流入速度 | 5 – 100 cm/s(應請求支持其它流速) |
壓差測量 | 0 – 1200 Pa |
介質測試區 | 100 cm² |
氣溶膠 | 灰塵(例如SAE灰塵),鹽(例如NaCl,KCl),液體氣溶膠(例如DEHS) |
氣溶膠濃度 | 對于未額外稀釋的SAE細塵(ISO A2 Fine),可達1,000mg/m3(ISO A2 Fine) |
壓縮空氣供應 | 6 – 8 bar |